Welcome Customer !

Membership

Shandong Drick Instruments Co.,LTD
Free Member ·1YRS
Other

Main Products: 试验机

Xwboo>Products

Shandong Drick Instruments Co.,LTDFree·1YRS

  • E-mail

    info@drktest.com

  • Phone

    0086-531-85868997,0086-531-85761369

  • Address

    3rd Building, A District,Shidai headquarters, No. 15 Lanxiang Road, Tianqiao District,Jinan City,Shandong Province, China. 250032

Contact Now

DRK-96 Ball Indentation Hardness Tester

Model
Nature of the Manufacturer
Other
Product Category
Plastic soft package testing instrument
Place of Origin
City, Province
Overview
DRK-96 Ball Indentation Hardness Tester
Product Details

Product details

The main technical parameters
1 six test load: 9.8N (pre-loaded), 49N, 132N, 358N, 612N, 961N
2 steel ball indenter: Φ5mm, Φ10mm
3 indentation depth indicator minimum scale value: 0.001mm
4 sample maximum allowable height:
10mm
5 Pressure head to wall distance:
55mm
6 Indication Accuracy: ±
1%
7 Timing Range: 10 ~ 90s, timing accuracy ± 0.1s
8 Effective measuring range: 0.150 ~ 0.350mm
9 Rack deformation: 0.05mm
10 Weight: 70kg
11 Equipment Dimensions (mm):
330 ×220 ×425

Host and accessories

Name

Specification

Quantity

Host

DRK-96

1

Precision sensors

LCD10KN

1

Deformation sensor

0-6.5mm

1

Indenter

Φ5mmΦ10mm

1

Base


1

Mini printer

WH-E22

1

Deformation collection system


1

Loading control system


1

Data micro-processing system


1

SCM software


1


 

Sample requirements
1 Samples should be of uniform thickness, smooth surface, no bubble, no mechanical damage and impurities.
2 Sample thickness of not less than 4mm, the sample size should ensure that each measuring point from the center and the edge of the sample is not less than 10mm, the distance between the centers of the measuring point is not less than 10mm. Recommended Sample size is 40 × 40 × 4mm or Ф50 × 4mm.

 

Note showing items
1 test must ensure that the force applied perpendicular to the sample surface, the specimen bearing surface shall be in good contact with the work surface during the test loading should be smooth, not be subjected to shock and vibration.
2 Installing or replacing the head table should first try a few times and the hardness in working condition, then hardness test can be carried out.
3 each test, to test three times.

Hardness maintenance
1 Hardness Tester must be installed in a dry place, there should be no vibration source nearby, no dust non-corrosive gases.
2 Use hardness timing, must be required hardness measurement range, if not determine the hardness of the test sample range, the load should be chosen to increase progressively from small to big.
3 Hardness chassis should be grounded well, cut off the power after use.
4 Hardness kept clean after use plus cover to prevent dust intrusion, to prevent rust.