DRK260 Respirator Resistance Tester
Product details
Introduction
The respirator resistance tester is used to measure the inspiratory resistance and expiratory resistance of respirators and respirator protectors under specified conditions.It is applicable to the national labor protection equipment inspection institutions, mask manufacturers for the general masks, dust masks, medical masks, anti-smog masks products of the relevant testing and inspection.
Standard
GB 19083-2010 Technical requirements for medical protective masks
GB 2626-2006 Respirator self-suction filter respirator against particulate matter
GB/T 32610-2016 Technical specifications for daily protective masks
NIOSH 42 CFR Part 84 Respiratory Protective Devices
EN149 Respiratory Protective Devices-Filtering half masks to protect against part
Features:
1. High-definition LCD display.
2. Digital differential pressure meter with high precision imported brand.
3, high precision imported brand of digital flowmeter,with high flow control accuracy features.
4. The respirator resistance tester can set up two modes: exhalation detection and inhalation detection.
5. Automatic pipeline switching device of respirator solves the problem of pipe extubation and wrong connection when testing.
6.Measure the exhalation resistance with the dummy head successively placed in 5 defined positions:
--facing directly ahead
--facing vertically upwards
--facing vertically downwards
--lying on the left side
--lying on the right side
Parameter
1. Flowmeter range :0 ~ 200L/min, accuracy is ±3%
2. Digital pressure difference meter range :0 ~ 2000Pa, accuracy: ±0.1%
3. Air compressor: 250L/min
4. Overall size :90*67*150cm
5.Test the inhalation resistance at 30L/Min and 95 L/Min continous flow
5. Power source :AC220V 50HZ 650W
6. Weight :55kg