Drk8023 melting point apparatus
Product details
Melting point measurement range: room temperature to 320℃.
Temperature display minimal indication: 0.1℃.Linear heating rate:0.2℃/min, 1℃/min,1.5℃/min,3℃/min,
tandard capillary dimensions: the outer diameter Φ1.2mm, inner diameter Φ1.0mm, height 120mm
The sample filling height: ≥3mm
Power supply: AC220V + 22V, 100W, 50Hz
Shown value error: when the environment temperature is 23℃±5℃, shall meet the following requirements: at less than 200 ℃ range: ±0.5 ℃
B) at 200 ℃ -320℃ range: ±0.8 ℃
Repetitive indication: when heating rate is 1.0℃/min, uncertain degree of samples is 0.3 ℃
Linear heating rate error: not larger than the nominal value of 10%
Instrument size: 390mm x 320mm x 240mm
Instrument net weight: 12kg