DRK8024A microscopic melting point meter
Product details
Melting point measurement range: room temperature -300℃
“The initial temperature" setting time: 50℃ -300℃ ≤6min 300℃ -50℃ ≤7min
Temperature display minimal indication: 0.1 C
Linear heating rate: 0.2 ℃/min,0.5℃/min,1℃/min,1.5℃/min,2℃/min,
3℃/min,4℃ /min,5℃/min eight file
Linear heating rate error: not larger than 10% of the nominal value
Shown value error: ≤200 ℃:±0.4 ℃ >200℃:±0.7 ℃
Repetitive indication: when the heating rate is 1.0℃/min, 0.3 ℃
Standard capillary dimensions: the outer diameter is Φ1.4mm inner diameterΦ1.0mm length 80mm
Sample: filling height :≥3mm
Communication interface: USB or RS232 by the selection keys
Power supply: AC220V + 22V, 100W, 50Hz
Instrument size: 365mm x 290mm x 176mm
Instrument net weight: 10kg