Product details
The main technical parameters:
Working temperature: room temperature to 300 ℃
Initial temperature setting rate: 50 ℃ to 300 ℃ is not more than 12 min
300 ℃ to 50 ℃ is not more than 15 min
Initial temperature setting accuracy: ± l℃
The minimum temperature reading: 0.1 ℃
Linear ramp rate: 0.2 ℃ / min, 0.5 ℃ / min, 1 ℃ / min, 1.5 ℃ / min, 2 ℃ / min,
3 ℃ / min, 4 ℃ / min, 5 ℃ / min eight gears
Linear heating rate deviation: ±10%
Jingyi heating temperature control accuracy: < 200℃±0.5℃
≥200℃±0.8℃
The thickness of the glass slide: 0.8 mm ~ 1.2 mm
Cover glass thickness: 0.17 mm
Microscope Working distance: Objective to sample> 7 mm Objective to table> 29mm
Hot jingyi size: 195 mm * 28 mm * 85 mm
Program-controlled digital display unit size: 400 mm * 130 mm * 400 mm
Instrument weight: 12.6 kg