GC-9802 Gas Chromatograph

Model
Nature of the Manufacturer
Other
Product Category
Solvent Residues Test
Place of Origin
City, Province
Overview
GC-9802 Gas Chromatograph
Product Details
Application 
To test the smell and solvent residues of printed packaging materials; test the quality and purity of the solvent. 

Working principle 
Sample of gas, liquid or solid is processed and passes through vaporizing chamber, brought to chromatographic column separation system. Carried by nitrogen, the sample mixture passes and contacts the chromatographic column, and the composition components of the mixture are separated by chromatographic column. Then every separated component is brought to the detector system, and transformed by the detector to equivalent electronic signal, which is collected and recorded as data, so as to make qualitative and quantitative analysis of the components of the sample mixture. 

1. Index of detectors

Thermal conductivity detector

Flame ionization detector

Electron capture detector

Flame photometric detector

(TCD)

(FID)

(ECD)

(FPD)

Sensitivity (limit of detection)

≥3000mv.mL/mg

(benzene)

≤1×10-11g/s

(hexadecane)

1×10-13g/ml(γ-666)

5×10-12g/s(P)

5×10-11g/s(S)

Baseline noise

≤10μV

≤1×10-13A

≤40μV

≤40μV

Baseline shift

≤60μv /30min

≤6×10-13A/30min

≤100μv/30min

≤100μv/30min

Linearity range

105

107

104

104

Settling time

≤2h

≤1h

≤2h

≤1h

 

2. Other parameters:

Temperature range

room temperature +10℃ ~ 400℃

Temperature controlling accuracy

±0.5%

Temperature resolution

0.1℃

Program period number

3 periods

Temperature rise speed

0~39℃/minute

Repeatability of temperature programming

≤1%

Column oven size

250mm × 250mm × 180mm

Instrument size

500mm × 540mm × 480mm

Instrument weight

35 kg

Power

1500W

QBT 2929, GB/T10004-2008, YBB


u Microcomputer temperature controller, four-way, digital display.

u Two kinds of commonly used detectors are optional at the same time.

u Can control temperature close to room temperature (room temperature +10℃).

u Column oven with large capacity has auto back opening door. Can perform three-period program heating.

u Optional configurations: packed column sample injector, capillary column sample injector, gas sample injector and so on.

u With fault self-diagnosis function, displaying fault location and fault nature at any time.

u Over temperature protection. When temperature of any way of the four ways exceeds set temperature, the instrument will stop heating automatically.

Main machine body, hydrogen flame ionization detector (FID), chromatographic column, data workstation, software, installation fittings, sample injector, headspace bottle, sealing plier 
Optional (for purchase): thermal conductivity detector (TCD), electronic capture detector (ECD), flame photometric detector (FPD), nitrogen and phosphor detector (NPD), chromatographic column. 