GC-9802 Gas Chromatograph
Application
To test the smell and solvent residues of printed packaging materials; test the quality and purity of the solvent.
Working principle
Sample of gas, liquid or solid is processed and passes through vaporizing chamber, brought to chromatographic column separation system. Carried by nitrogen, the sample mixture passes and contacts the chromatographic column, and the composition components of the mixture are separated by chromatographic column. Then every separated component is brought to the detector system, and transformed by the detector to equivalent electronic signal, which is collected and recorded as data, so as to make qualitative and quantitative analysis of the components of the sample mixture.
1. Index of detectors
| Thermal conductivity detector | Flame ionization detector | Electron capture detector | Flame photometric detector |
(TCD) | (FID) | (ECD) | (FPD) |
Sensitivity (limit of detection) | ≥3000mv.mL/mg (benzene) | ≤1×10-11g/s (hexadecane) | 1×10-13g/ml(γ-666) | 5×10-12g/s(P) 5×10-11g/s(S) |
Baseline noise | ≤10μV | ≤1×10-13A | ≤40μV | ≤40μV |
Baseline shift | ≤60μv /30min | ≤6×10-13A/30min | ≤100μv/30min | ≤100μv/30min |
Linearity range | 105 | 107 | 104 | 104 |
Settling time | ≤2h | ≤1h | ≤2h | ≤1h |
2. Other parameters:
Temperature range | room temperature +10℃ ~ 400℃ |
Temperature controlling accuracy | ±0.5% |
Temperature resolution | 0.1℃ |
Program period number | 3 periods |
Temperature rise speed | 0~39℃/minute |
Repeatability of temperature programming | ≤1% |
Column oven size | 250mm × 250mm × 180mm |
Instrument size | 500mm × 540mm × 480mm |
Instrument weight | 35 kg |
Power | 1500W |
QBT 2929, GB/T10004-2008, YBB
u Microcomputer temperature controller, four-way, digital display.
u Two kinds of commonly used detectors are optional at the same time.
u Can control temperature close to room temperature (room temperature +10℃).
u Column oven with large capacity has auto back opening door. Can perform three-period program heating.
u Optional configurations: packed column sample injector, capillary column sample injector, gas sample injector and so on.
u With fault self-diagnosis function, displaying fault location and fault nature at any time.
u Over temperature protection. When temperature of any way of the four ways exceeds set temperature, the instrument will stop heating automatically.
Main machine body, hydrogen flame ionization detector (FID), chromatographic column, data workstation, software, installation fittings, sample injector, headspace bottle, sealing plier
Optional (for purchase): thermal conductivity detector (TCD), electronic capture detector (ECD), flame photometric detector (FPD), nitrogen and phosphor detector (NPD), chromatographic column.