LPS filter cartridges are integrity testable membrane filters specially developed for sterile filtration of Large Volume Parenterals (LVP) and Small Volume Parenterals (SVP). They are characterized by a unique asymmetric, hydrophilic polyethersulfone membrane, offering high throughputs, low protein adsorption and superior microbiological safety. These filters add the sterility assurance for parenterals.Features and Benefits● Low protein binding ● Absolute bacteria retention● Long service life and high flow rates Qu......
LPS filter cartridges are integrity testable membrane filters specially developed for sterile filtration of Large Volume Parenterals (LVP) and Small Volume Parenterals (SVP). They are characterized by a unique asymmetric, hydrophilic polyethersulfone membrane, offering high throughputs, low protein adsorption and superior microbiological safety. These filters add the sterility assurance for parenterals.Features and Benefits● Low protein binding ● Absolute bacteria retention● Long service life and high flow rates Quality● Cartridges produced in a controlled environment● Manufactured according to ISO9001 certified Quality Management System
Specification
LPS Filter Cartridge for Parenterals
Materials of Construction |
Membrane | Single Layer Hydrophilic PES |
Support | PP |
Core, Cage and Drainage | PP |
End Caps | PP (222/226 with encapsulated stainless steel reinforcing ring) |
O-rings/Gasket | Silicone/EPDM/FKM |
Sealing technology | Thermal Bonding, No Adhesives |
Dimensions |
Diameter | Φ68 |
Length | 5 inch, 10 inch, 20 inch, 30 inch, 40 inch |
Filtration Area, ft2 |
≥ 5.7 per 10-inch cartridges |
Pore Size, μm |
0.1, 0.22, 0.45, 0.65, 0.8, 1.2, 3.0 |
Maximum Differential Pressure |
Forward: 4.2 bar @ 23 ℃; 1.5 bar @ 85 ℃ |
Integrity Test-water bubble point at 23 ℃ |
0.1 μm | ≥ 3800 mbar (58 psig) |
0.22 μm | ≥ 3400 mbar (49 psig) |
0.45 μm | ≥ 2300 mbar (33 psig) |
0.65 μm | ≥ 1300 mbar (19 psig) |
0.8 μm | ≥ 600 mbar (8 psig) |
1.2 μm | ≥ 400 mbar (6 psig) |
Bacterial Retention (for 0.1 μm & 0.22 μm only) |
Passed the bacterial challenge testing using Brevundimonas diminuta(ATCC19146) at a minimum challenge concentration of 1 x 107 CFU/cm2. |
Extractables, per 10-inch cartridge |
≤ 50 mg after 24-h soak in water |
Effluent Particle Level |
Meets the requirements of Chinese Pharmacopoeia 2010, Volume Ⅱ, appendix Ⅸ C. |
Non-Fiber Releasing |
Meets the criteria for a “non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6). |
Toxicity |
Component materials meet GB/T 14233.2<2005>of Chinese National Standard for Safety Tests. |
TOC |
< 500 ppb after a water flush of 60 liters per 10-inch cartridge |
Multiple Sterilization Cycles |
15 steam-in-place sterilization or autoclave cycles of 30 min at 123℃ |
Oxidizable Substances |
Meets the criteria of Chinese Pharmacopoeia 2010, volume Ⅱ for purified water. |
Bacterial Endotoxins |
< 0.25 EU/ml as determined by the LAL test |