LPM filter cartridges are characterized by high-performance, hydrophilic mixed cellulose esters membrane, offering high flow rates, low protein adsorption and superior microbiological safety. These filters are ideal for prefiltration for protecting sterilizing filters or chromatography column prior to downstream purification in plasma fractionation production.Features and Benefits● Low protein binding● Superior throughputs and cost-effective● High retention efficiency for more critical degrees of prefiltrationQuali......
LPM filter cartridges are characterized by high-performance, hydrophilic mixed cellulose esters membrane, offering high flow rates, low protein adsorption and superior microbiological safety. These filters are ideal for prefiltration for protecting sterilizing filters or chromatography column prior to downstream purification in plasma fractionation production.Features and Benefits● Low protein binding● Superior throughputs and cost-effective● High retention efficiency for more critical degrees of prefiltrationQuality● Cartridges produced in a controlled environment● Manufactured according to ISO9001 certified Quality Management System
Specification
LPM Filter Cartridge for Plasma Fractionation
Materials of Construction |
Membrane | Hydrophilic MCE |
Support | PP |
Core, Cage and Drainage | PP |
End Caps | PP (222/226 with encapsulated stainless steel reinforcing ring) |
O-rings/Gasket | Silicone/EPDM/FKM |
Sealing Technology | Thermal Bonding, No Adhesives |
Dimensions |
Diameter | Φ68 |
Length | 5 inch, 10 inch, 20 inch, 30 inch, 40 inch |
Filtration Area, ft2 |
≥ 6.4 per 10-inch element |
Pore Size, μm |
0.1, 0.22, 0.45, 0.65, 1.0, 3.0, 5.0 |
Maximum Differential Pressure |
Forward: 4.2 bar @ 23 ℃; 1.5 bar @ 85 ℃ |
Extractables, per 10-inch cartridge |
≤ 50 mg after 24-h soak in water |
Effluent Particle Level |
Meets the requirements of Chinese Pharmacopoeia 2010, Volume Ⅱ, appendix Ⅸ C. |
Non-Fiber Releasing |
Meets the criteria for a “non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6). |
Toxicity |
Component materials meet GB/T 14233.2<2005>of Chinese National Standard for Safety Tests. |
TOC |
< 500 ppb after a water flush of 60 liters per 10-inch cartridge |
Multiple Sterilization Cycles |
15 steam-in-place sterilization or autoclave cycles of 30 min at 123℃ |
Oxidizable Substances |
Meets the criteria of Chinese Pharmacopoeia 2010, volume Ⅱ for purified water. |
Bacterial Endotoxins |
< 0.25 EU/ml as determined by the LAL test |