BT100-1F+DG (0.0002-48ml/min)

Product model
BT100-1F Dispensing type Peristaltic Pump
[Product introduction]
Flow range: 0.0002-48ml/min
The best filling volume: 0.1-100mL
Various pump head for optional, DG series, YZ series、BZ series
Working mode: Flow mode, Dispensing mode
Flow rate rectify, Dispensing volume rectify function
Flow accuracy within ±0.4%
[Application]
Lab, filling
Reagent packing, culture medium filling
[Technical parameter]
Model | BT100-1F Dispensing type Peristaltic Pump |
CPU | ARM Cortex-0 Core,STM32 Chip |
Motor drive chip | SANYO THB7128 drive chip |
Speed resolution | 0.1RPM |
Distribute function | liquid Quantitative filling accurately |
Dispensing type range | 0.1mL-999L |
Filling times | 1-9999 times |
Interval time | 0.1S-24Hr |
Display function | 128×32 Dot matrix LCD display, flow rate and speed display in the same screen |
Direction control | Clock-wise/counter clock-wise direction reversible |
Operation mode | Membrane keypad adjust the speed, also support external signal adjust speed |
Full speed function | Full speed button to realize fast filling and drain |
Power-down memory function | After power on, running as before |
Back suck function | Running under Dispensing type mode, pump reverse a certain angle after stopping, to prevent liquid drip |
Calibration function | Flow rate rectify function, accuracy ±0.5% |
Dispensing type Calibration function | Input filling volume , it rectify the distributed liquid automatically; accuracy ±0.5% |
External control function | Different model external control module for Start/Stop, direction, flow rate; 0-5V、0-10V、4-20mA、0-10KHz 4 module for option |
Communication function | Optional communication module, RS485 communication control (via Modbus protocol) |
Foot switch control | Optional |
Bracket | Optional (for installing the nozzle) |
Nozzle | Various specification nozzle for option, nozzle inner diameter 2mm、2.5mm、3mm、4mm、6mm |
Power consumption | 35W |
Power supply | Wide AC90-260V,50Hz/60Hz |
Operation condition | Temperature 0 to 40 °C, Relative humidity < 80% |
Driver weight | 4.5KG |
IP rate | IP31 |
[Selection guide]
Filling volume | Filling time (Sec) | Accuracy | Recommend driver | Recommend pump head | Recommend tube |
50uL | ≥1.18 | ±3.5% | BT100-1F | DG-1A | 0.38×0.85 |
100uL | ≥0.46 | ±2.0% | 1×1 |
200uL | ≥0.93 | ±1.2% | 1×1 |
0.5mL | ≥0.85 | ±1.2% | 2×1 |
1mL | ≥0.80 | ±1.0% | BT100-1F | YZ15 | 14# |
2 mL | ≥0.64 | ±1.0% | 16# |
3 mL | ≥0.97 | ±1.0% | 16# |
5 mL | ≥0.82 | ±1.0% | 25# |
Driver | Pump head | Roller of the pump head | Channel QTY | Tube | Speed range (RPM) | Flow range (mL/min) |
BT100-1F | DG-1(2,4,6)A | 6 | 1,2,4,6 | 0.6×1 | 0.1-100 | 0.003-3.27 |
1×1 | 0.007-6.45 |
2×1 | 0.02-20.60 |
3×1 | 0.04-37.5 |
DG-1(2,4,6)B | 10 | 1,2,4,6 | 0.6×1 | 0.1-100 | 0.0026-2.64 |
1×1 | 0.006-5.96 |
2×1 | 0.017-16.19 |
3×1 | 0.028-27.65 |
BT100-1F | YZ15 | 3 | 1,2 | 13# | 0.1-100 | 0.008-7.45 |
14# | 0.027-29.87 |
19# | 0.052-55.7 |
16# | 0.086-180.4 |
25# | 0.18-186.6 |
17# | 0.32-315.0 |
18# | 0.41-460.0 |
YZ25 | 3 | 1,2 | 15# | 0.1-100 | 0.17-175.0 |
24# | 0.29-350.0 |
BZ15-14 | 3 | 1,2 | 14# | 0.1-100 | 0.023-24.5 |
BZ15-16 | 16# | 0.079-78.5 |
BZ25A | 24# | 0.25-278.4 |
Remarks:
Above data are tested actually by silicone tube, various import tube for option
Import tube to prolong tube lifetime and accuracy
[Dimension] 