KM 50L Particle Counter ACS Plus 328

Model
Nature of the Manufacturer
Other
Product Category
Particle counters
Place of Origin
City, Province
Overview
KM 50L Particle Counter ACS Plus 328
Product Details
Particle Counter ACS Plus 328   0.3UM  /  0.1um size 
Features:


• Size range from 0.1 um with a maximum of 16 freely selectable particle size channels
• Flow rate 1 CFM, optionally with 50 L/ min.
• Built-in measuring data memory for 50000 measurements
• Access authorization concept with password protection
• Software PAS for evaluation of the measured data via PC
• 8.4“ LCD TFT display
• Optionally with removable/ rechargeable Li-Ion batteries for easy operation at various measuring points
• CAN-BUS Interface for management of a network spanning up to 32 particle sensors (slaves) without need of additional hardware
• Weight: approx. 10 Kg
• Made in Germany by KM OptoElektronik GmbH

Easy handling

The large-size display with touch-screen function and concise menu navigation ensures a simple and trouble-free implementation even of major settings. Furthermore, the built-in access authorization concept offers an easy way to grant specific access permissions for individual operators. In addition the non-reflecting display surface allows a data reading from greater distance.

Flexible monitoring

Adapted to application requirements the high-end counter ACS Plus may be used as a stand-alone unit or as a network monitoring system. Via device interface with appropriate sensors it is always possible to monitor and record temperature patterns, relative humidity, differential pressure and volume flow, if desired. The implementation of an alarm panel has also been considered.

GMP conforming Design

One of the main conditions for the application of equipment within a GMP controlled area is a surface design being suitable for the pharmaceutical industry. This of course also applies to particle counters. In this regard the capabilities for clean room applications as well as the resistance to detergents and disinfectants are most important technical features which have been implemented with the design of the surface characteristics.
 
Specification
 

ModelACS Plus 328
Light SourceLaser diode
Size Range(um)0.3 – 10(25)
Concentration Limit1 Mio./ 1 CFM
Flow Rate28.3L/min (50L/min optional)
Weight (kg)10
Dimension L*W*H(mm)410x305x140
Flow-GeneratorVacuum pump
Flow-ControlAutomatically control
CalibrationPer ASTM F- 328 ASTM F- 649; Latex-Aerosol
Display and OperationRobust and bright 8.4“ LCD TFT display SVGA, 800x600 pixel, luminous intensity 350cd/ m²,sunlight readable transmissive TFT active-matrix screens with cold cathode backlighting (CCFL),wide viewing angle
Sampling/Rinsing Time1s to 99h,59min,59s
Sampling ModesSingle measuring with 1 - 9 measurements for each cycle and automatic averaging, or continuous measuring with automatic averaging of 2 - 9 measurements each
StandardISO 14644-1, VDI 2083
PrinterThermal printer 58 mm
Communication ModeStandard RS 232 for single application, RS 485 for networking
Data Content per
Record
Date, time, ID number, alphanumeric key, measuring and rinsing time,16 particle size channels with cumulative particle count
Audible AlarmSingle particle detection by acoustics, threshold exceeding alarm adjustable