Battery Charge And Discharge Tester

Wuhan Goldhome Hipot Electrical Co., Ltd.

Model
Nature of the Manufacturer
Manufacturer
Product Category
Other Testing Machines
Place of Origin
HuBei, WuHan
The tester can be used as discharge load in the off-line state of battery, and the constant current discharge of the set value can be realized by continuously regulating the discharge current. During discharge, the instrument will automatically stop the discharge when the battery terminal...

Introduction


The tester can be used as discharge load in the off-line state of battery, and the constant current discharge of the set value can be realized by continuously regulating the discharge current. During discharge, the instrument will automatically stop the discharge when the battery terminal voltage, or the monomer voltage, falls to the set lower limit, or the set discharge time, or the set discharge capacity, and record all valuable, continuous process real-time data

Main Parameter


Battery

DC10-300V

Charge/discharge voltage

DC10-300V

Charge/discharge current

Charge 40 A 80 A/120A/80A

work pattern

Single machine mode, parallel host mode, parallel slave mode, remote controlled mode

Protection performance

l Battery test voltage overvoltage protection,Undervoltage downtime, overcurrent protection,Reverse protection,65℃ Protection against overheating,And LCD tips,buzzer alarm

Single voltage acquisition

communication distance can reach more than 100 meters, compatible with 2 V/4V/6V/12V single cell voltage monitoring ,1~25 groups of radio frequency wireless monitoring module, up to 300 batteries, A single wireless monitoring module can monitor 12 single cells

control accuracy

discharge current ≤±1%;Group terminal voltage≤±0.5%;Single voltage:≤±0.1%

PC communications

RS485,USB

Data storage capacity

Built-in SD card 8 G capacity, transfer U disk 16 capacity

working power supply

Voltage

Working power: single phase AC220V (–20%~30%), frequency :45~65 Hz;

Charging Power: Reference nameplate parameter or chassis identification

high-voltage insulation test

Input - housing:2200Vdc 1min

Input-output:2200Vdc 1min

Output - housing:700Vdc 1min


details



Battery measurement principle


Due to the complexity of the electrochemical reaction of batteries, as well as the differences in various materials, structures, manufacturing processes and use environments, the characteristics of batteries from different manufacturers are quite different. Even batteries produced by the same manufacturer will have certain characteristics of their monomers. The discreteness. So far, there is no simple and effective method in the world that can quickly and accurately determine the battery performance. The detection and failure prediction of battery performance is still a very complicated electrochemical measurement problem.

Fixed acid-isolated explosion-proof lead-acid batteries that have been widely used in power, communications, finance, transportation and other industries can understand the battery status by measuring terminal voltage, checking electrolyte density, liquid level, and temperature. However, the sealed and lean-liquid design of valve-regulated lead-acid batteries makes it difficult for us to grasp its health status. The detection and maintenance methods for acid-isolated and explosion-proof batteries are no longer suitable for valve-regulated batteries. This is the current battery operation management The shortcomings and difficulties.

At present, the commonly used detection methods are to measure the terminal voltage of the battery at ordinary times and to conduct a check discharge capacity test every year. We believe that:

1. There is no corresponding relationship between the terminal voltage and the capacity of the battery in the floating state.

We know that even a battery with poor performance may measure a qualified voltage in a floating charge state. Therefore, the terminal voltage in the floating charge state cannot truly reflect the battery performance.

2. The full capacity discharge test is still the most accurate and effective method to test the actual capacity of the battery pack.

We know that the capacity of a battery pack is equal to the capacity of the battery with the worst performance in the battery pack. Therefore, the detection of the battery pack can be transformed into the detection of the backward battery, and the capacity of the battery pack can be obtained by finding the backward battery and measuring the capacity of the battery.

Discharge the battery pack with a prescribed constant current, and monitor the voltage of each battery at the same time. When the voltage of any one of the batteries drops to the end voltage, the discharged capacity is the actual capacity of the battery pack.


  Wuhan Goldhome Hi pot Electrical Co.,ltd(short for Goldhome Hi pot) was established in 2008, its predecessor was state-owned third industry enterprises, one of the earliest enterprises specially in develop, design and manufacture high voltage testing equipment in China.In 2016, merged with Wuhan Sang Electric Co., Ltd. , we realize advantageous compartmentalize and resources reorganization.Now it has become a high-tech enterprise integrating R&D, production, marketing installation and commissioning.1.our-historyOur FactoryGoldhome Hi pot is located in Huading Mould Industrial Park, Economic and Technical Development Zone, Wuhan City, Hubei Province, China, working area more than 4000 square meters.As a high-tech enterprise, we are supported by natural advantages from various research universities in Wuhan city. Goldhome Hi pot works closely with professors from State Grid Power Science Research Institute, Electrical Engineering Colleges of Wuhan University and other scientific research institutions, constantly improving technology and updating products . Now Goldhome Hi pot has more than 20 patented technologies, and many software Copyrights!2.our factory-1product and product applicationOur ProductOur advantage products included AC Resonant Test System, PD Free Test System, AC/DC Hi pot Tester, Primary Current Injection, VLF Hi pot Tester, transformer tester,impulse generator,Transformer oil tester,HV Switch gear Tester ,Insulation Tester,ground Resistance Tester,cable fault locator,relay test set,CT PT analyse,arr ester tester and sf6 gas analyzer,etc.Product ApplicationAll the performance indexes of above testing equipment can reach the leading level in China, widely used in power supply bureau, power plant, power construction, railway, petroleum, chemical industry, metallurgy, coal and other industries.In recent years,our factory continues to expand overseas markets, some competitive products are already exported to American and European countries, Africa, Middle East, etc , and well praised by customers.product applicationOur CertificateAll product process is strictly in accordance with international standards and power industry regulations. And the products are through the national, provincial power authorities testing, as well as CE certification! At present,0ur company has received ISO9001:2015 quality management system certification, ISO14001:2018 occupational health and safety management system certification, and ISO14001:2015 environmental management system certification,and also had got SGS certification.We has more than 20 patented technologies,such as The utility model patent of a frequency conversion series resonance test device,The utility model patent of Lightning impulse voltage testing equipment,The utility model patent of DC high voltage generator,The utility model patent of microcomputer relay protection tester,The utility model patent of High voltage switching characteristic tester., etc.4.certificateoverseas market & our serviceOverseas MarketOur products are sold to Europe, Asia, Africa, North and South America and Oceania, Middle East, etc.and are well received by customers.We had won the bids for the purchase of high-voltage no-partial discharge test equipment from the Polish state-owned insulator production plant, the Myanmar National Grid cable fault tester purchase project, and the high-voltage test equipment purchase project of a Russian power installation and test engineering company.We also sell to end customers, and a single device can also be sold. For example, there are power maintenance and installation companies in Indonesia, South Africa, Denmark, and South Korea, and universities and research institutions purchase and use our test equipment.At the same time, our sales targets also include agencies. At present, we already have cooperated with agents in Singapore, Israel, Mauritius , etc.We insist on providing customers with first-class products and first-class services with reliable product quality and perfect service team.5market-1Our service1.Strict factory testAll products will be assembled and test strictly accord with our product process standard. All qualified products will be issued factory release reports and certificate labels.Defect products or non-qualified products won’t be allowed to make deliver.2.Providing technical materialsDuring the product acceptance, we provide the related manual sheets, packing list,qualified certificate labels, factory test reports and other technical documents.3. Support technical guidanceWe also provide the training for the operators. Complete training includes the skilled operations of our equipment, maintenance and troubleshooting . it also ranges the commissioning and test of our products at site!After-sale service1.First-use responsibility systemAfter the product is delivered, the technician can guide the operation at site if the client requires, and be responsible for the first use of each part2.One year warrantyAll products warranty is 1year, life-time maintenance .During the warranty period, if any defects or damaged because of product’s quality, maintenance or replacement can be provided with free charge in order to let our client’s satisfied !3.TroubleshootingIf the troubleshooting is not difficult, and client can handle with it by themselves, we can offer the call-on line guide within 1 hour!If the troubleshooting is difficult to handle with for client, we can promise that the best solution can be provided within 24 hours!