Digital Variable Area Flow Meter
The Digital LCD Display Variable Area Flow Meter is special configured base on the mechanical type variable area flow meter.the flow meter have LCD display and totaliser to show the current flow rate and the accumulate flow amount on the screen.The Digital LCD Display Variable Area Flow Meter can output the 4-20mA signal and the switch signal for industrial system control.The digital variable area flow meter is widely used in the flow rate monitoring of liquid and gas.The transmitter can be explosion proof processed for some easy explosion media.


| Measure Medium | Water,Oil,Compressed Air,Nitrogen |
| Flow range | Water 2.5 l/h to 200 m3/h |
| Air 0.07 to 4500m3/h (201.013bar) | |
| Flow span | 10:01 |
| Accuracy | Standard accuracy: 1.5% High accuracy: 1.0% |
| Standard accuracy (PTFE-Liner): 2.5% High accuracy (PTFE-Liner): 1.5% | |
| Connections | Flange connection |
| Standard: DIN2501DN15-DN200 ) | |
| Special: DIN2512DN15-DN200 | |
| ANSI1/2-8 | |
| The GB JIS users needs | |
| Threaded fittings DIN11851SC40-SC80[1-1/2-3] | |
| According to demands of users | |
| Steam Jacket Connections | Internal threads 1/2 |
| NPT | |
| Flange connection DN15 PN40 DIN2501 | |
| According to demands of users | |
| Max.Allow. pressure | 250bar |
| Temperature range | Fluid temperature |
| -40 to+200Standard type | |
| -20 to+125PTFE-Liner type | |
| -40 to+400High temperature type | |
| Ambient temperature | |
| -40 to+70 | |
| Materials | SS304,SS304L,SS316,SS316L |
| PTFE-Liner | |
| Hastelloy C | |
| Ti | |
| According to demands of users | |
| Protection class | IP67 |
| Supply power | 12-28V DCtwo-wire systembattery power |
| Ex-Design | ExdIIBT6 |
Common faults of Digital Variable Area Flow Meter and their solutions
1. No current output of Digital Variable Area Flow Meter
1)First, check whether the wiring is correct.
2)Whether the LCD has a display, if there is an output, the output tube is mostly broken, and the circuit board needs to be replaced.
3) Missing calibration value. Due to E2PROM fault, the instrument calibration data is lost, and there will be no output current, and the current will remain unchanged. Solution: available data recovery operation. If it doesn't work, you can first set the data in password 2000, and then set the data in password 4011. The method is to calibrate the data from RP to 99% with hand push pointer.
2. Digital Variable Area Flow Meter without field display
1)Check whether the wiring is correct.
2) Check whether the power supply is correct.
3) Reinstall the LCD module and check whether the contact is real.
4) For multi wire power supply mode, check whether terminals 12 and 13 are connected with ammeter or short circuit.
3. Liquid crystal display 0 or full scale on site for Digital Variable Area Flow Meter
1) Check the range and zero point parameters set in the 2000 password. Zero is required to be less than span, and the two values cannot be equal.
2) Check whether the sampling data is up, and use the hand push pointer to see the change of the sampling value. If there is no change, it is generally the circuit board sampling circuit fault, and the circuit board needs to be replaced.