Welcome Customer !

Membership

XwbooInstrumentationElemental AnalyzerOther Elemental AnalyzersHT-TA261 Nitrogen/ Oxygen Transmitter
  • $item.Title
  • $item.Title

HT-TA261 Nitrogen/ Oxygen Transmitter

Jiuyin Science and Technology Chengdu Co., Ltd.

Model
Nature of the Manufacturer
Manufacturer
Product Category
Other Elemental Analyzers
Place of Origin
SiChuan, ChengDou
HT-TA261 Nitrogen/ Oxygen Transmitter

    Measurement Principle: Zirconia

    Display Mode: 5 dight high-light LED display, external display module available (optional)

    Measurement Range: 10/100/1000PPm~21.000% O2/N2

    Measurement Accuracy:10PPm~21.00%≤±1%FS

    10PPm~1.00%≤±2%FS

    10PPm~100PPm≤±5%FS

    Repeatability: ≤±1.0%FS

    Response Time: T90<30S

    External Port: 4-20mA.DC (Non-isolated output, load resistance<1KΩ)

    1 way of programmable alarm output (Dry contact type, maximum load 220VAC/2A)

    Display module port

    Communication Mode: RS485

    Power Supply: DC24V±10%,1.5A

    Environment Temperature: -10℃~+50℃

    Storage Environment Humidity: <90%RH, non-condensing

    Operation Environment Humidity: <99%RH, non-condensing

    Sample Gas Temperature: 0℃~60℃

    Sampling Mode: Gas in

    Gas Inlet Pressure: Stable pressure (control the pressure within the recommended limits to make sure the flow is within 400~600mL/min)

    Gas Outlet Pressure: Free venting (under safe conditions)

    Background Gas: Mixed gas of nitrogen and oxygen

    Sensor Service Life: >5 years (under normal conditions of use)

    Gas Circuit Port: Φ6 plastic tube fitting

    Installation Method: Instrument: embedded mounting

    Oxygen sensor: fixing clamp



    a. The instrument is small and easy to maintain;

    b. The imported new-type zirconia sensor has a long service life, high sensitivity and rapid response speed;

    c. There are N2 and O2 types of display mode, and N2 displays in (N2%=100-O2%);

    d. Automatic temperature compensation could eliminate the impact of environment temperature changes on the measured value;

    e. Upper and lower alarm limit can be set among the full range;

    f. Measurement range is 10PPm~21.000% O2;

    g. The instrument allows multi-point fractional liner calibration, meeting the need for an accurate measurement of gas concentration among the full range.

    h. Built-in standard RS485 communication port allows a unidirectional or bidirectional communication with computer or other digital communication equipment.



    Air separation industry: air separation fractionation, nitrogen generator, gas purification;

    Chemical/petrochemical industry: chemical production, high purity gas production;

    Electronics: reflow soldering, wave soldering, soldering powder production, semiconductor processing furnace;

    Metal: heat treatment/low temperature annealing, steel, alloy or powder metal production;

    Regular application: environmental monitoring, glove box;

    All kinds of purification equipment, chemical process, glove box, protective gas furnace and other online analysis of oxygen concentration measurement.



    Measurement range

    Measured gas pressure

    Background gas composition



    Jiuyin Technology Chengdu Co., Ltd. is a high-tech enterprise with more than 10 years of experience in the gas analysis industry, specializing in the research, development, production and sales of gas analysis instruments and systems, with a certain scale and comprehensive strength. Our company has always adhered to the development strategy of innovation and fully relied on the advantages of human resources to provide customers with products with high efficiency, high reliability and high cost performance. Since its establishment, the company has given full play to its own advantages, perfected its sales network, paid attention to its corporate image, followed the management concept of " technology forerunner", actively adopted international advanced technology, and possessed stable and excellent suppliers, partners and customer resources, which can provide customers with high performance-price ratio analysis instruments, analysis modules, pretreatment components, gas analysis systems and other services. The main products are: oxygen analyzer, dew point meter, humidity meter,gas chromatograph and other 9 categories, 16 series and 100 balance products, which are widely used in scientific research, environmental protection, energy-saving transformation, metallurgy, building materials, chemical industry, petroleum, power plant, pharmaceutical and food industries and other fields. The company has obtained ISO9001 quality management system certification, 20+ national patents and 20+ computer software copyright certificates, and was certified as a high-tech enterprise in September 2018, continuously developing high-tech products suitable for customers' site with outstanding service quality and continuously innovative technical service strength. The company now has an experienced technical support team to provide users with 24 / 7 technical support to solve the user's on-site debugging and use problems as soon as possible. Jiuyin Technology Chengdu Co., Ltd. sincerely welcomes friends from all walks of life at home and abroad to come to negotiate business and patronize guidance.
    Similar Product Recommend