HZBS-X3 Closed Cup Flashpoint Measuring Instrument Manual Closed Flash Point And Fire Point Tester

Model
Nature of the Manufacturer
Manufacturer
Product Category
Flash Point Tester
Place of Origin
City, Province
Overview
HZBS-X3 Closed Cup Flashpoint Measuring Instrument Manual Closed Flash Point And Fire Point Tester
Product Details

III.Instrument Features

With touch screen to replace the keyboard operation, is used to measure the flash point value of petroleum products. Using advanced technology abroad, LCD screen full English display man-machine dialogue interface, full screen touch keys, convenient and fast, open beam, fuzzy control integrated software, modular structure.

n Standard modular design, one main engine can control the number of open or closed test furnaces, be freely exchanged between the test furnace, According to user needs, greatly improving the efficiency and economic benefits.

n High-speed signal processor controlled, high-precision analog / digital converter circuit, the instrument is reliable and high precision.

n Auto-ignition.

n Automatic measurement of atmospheric pressure and automatic correction.

n Adaptive PID control algorithm to automatically adjust the temperature according to standard curve.

n The whole experiment process is completed automatically, including detection, opening the cover, ignition, alarm, cooling and printing.

n Automatically stop detection and alarm when the temperature is out of range.

n Thermal micro-printer, making printing more efficient and more beautiful, with off-line printing;

n History records with time stamp, up to 256

n Large screen with 320x240 graphical LCD display, showing a rich, full character display.

n touch screen operation, friendly interface; The screen saver function, extend the life of the LCD;

n Interface, RS232 interface for printer or downloads on PC

IV.Technical Parameters

1. Temperature Measurement

Ø Temperature measuring range: -59.9°C~299.9°C

Ø Temperature Control Range: Room temperature~300°C

Ø Repeatability: 0.025X(x=twice successive test average value)

Ø Resolution: 0.1°C

Ø Precision: 0.5%

Ø Temperature Sensor: Platinum Resistor(PT100)

Ø Flash Fire Sensor: Ion detection ring

2. Ambient Temperature: 0 ~ 40°C

3. Relative Humidity: ≤ 85% RH

4. Power Supply: AC220V ± 20% ; 50HZ ± 10%

5. Power Consumption: approx 400W

6. Overall Dimension: 440X280X290(mm)