Welcome Customer !

Membership

Fuzhou Fuguang Water Science&Technology Co.,Ltd
Free Member ·1YRS
FuZhou,FuJian
Manufacturer

Main Products: Ammonia Nitrogen Analyzer|Chlorine Dioxide Analyzer/Residual Chlorine Analyzer|Hardness Tester Ammonia Nitrogen Analyzer|Chlorine Dioxide Analyzer/Residual Chlorine Analyzer|Hardness Tester

Xwboo>Products

Fuzhou Fuguang Water Science&Technology Co.,LtdFree·1YRS

  • E-mail

    fgwater@fuguang.com

  • Phone

    400-668-8627

  • Address

    ADD:Building B, Chenxing Science and Technology Park, No. 168, Jinyan Road, Cangshan District, Fuzhou City, Fujian Province

Contact Now

9185 sc ozone analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
Ozone analyzer
Place of Origin
City, Province
Overview
9185 sc ozone analyzer
Product Details
typical application
The ozone analyzer is suitable for ozone disinfection monitoring in drinking water, sewage, industrial process water and pure water industries.
 
Instrument characteristics
● Ozone analyzer adopts selective membrane electrode;
● No interference from PH, chlorine, bromine, chlorine dioxide or hydrogen peroxide in the sample;
● The process of replacing the membrane takes only a few seconds;
 
Detection principle
The detection principle of the ozone analyzer: O3 is reduced on the gold electrode (cathode): O3 + H2O + 2e-→ O2 + 2OH-
On the silver reference electrode (anode): silver is oxidized to: Ag +, reacts with bromide ion in the electrolyte: 2Br- + 2Ag → 2AgBr + 2e-
The amount of reduction of O3 on the cathode produces a current proportional to the concentration of O3.
Scan the QR code to read on your phone
Technical index
Measuring range5ppb~20 ppm 
Accuracy± 3% or ± 10ppb, whichever is greater
Sample flow14 L/hr
Lower detection limit5 ppb
Response time90% ≤90 sec
Sample flow200~250 mL/min
Ambient temperature0~45°C
Sample temperature2~45°C
Power consumption1.5 W
Calibration intervalFebruary
Maintenance interval6 months, typical maintenance is membrane replacement, electrolyte;
Installation methodPanel installation;
Injection connection1/4-in. O.D.
Drain connection1/2-in. I.D.
Alarm OutputIP 66 (NEMA 4X)

 

Similar Product Recommend