SUP-DO7013 Electrochemical dissolved oxygen sensor

Model
Nature of the Manufacturer
Manufacturer
Product Category
Place of Origin
City, Province
Overview
SUP-DO7013 Electrochemical dissolved oxygen sensor is widely used in Aquaculture, water quality testing, informational data collection, IoT water quality testing etc.
Product Details
  • Specification

Measurement

DO value in water

Measure range

0~40.00mg/l

Resolution

0.01mg/l

Temperature range

-20~60°C

Type of sensor

Galvanic Cell sensor

Measuring accuracy

<0.5mg/l

Output mode

RS485 port*1

Communication protocol

Compatible with standard MODBUS-RTU protocol

Communication mode

RS485  9600,8,1,N  (by default)

ID

1~255 Default ID 01 (0x01)

Fixing method

RS485 remote setting calibration and parameters

Power supply mode

12VDC

Power consumption

30mA @12VDC

 

  • Introduction

Electrochemical dissolved oxygen sensor

  • Intelligent module communication protocol Introduction

Communication port: RS485

Port setting: 9600,N,8,1 (by default)

Device address: 0x01 (by default)

Protocol specifications: Modbus RTU

Commands support: 0x03 read register  

0X06 write register| 0x10 continuous write register

 

  • Information frame format

0x03 read data [HEX]

01

03

×× ××

×× ××

×× ××

Address

Function code

Data head address

Data length

Check code

0x06 write data [HEX]

01

06

×× ××

×× ××

×× ××

Address

Function code

Data address

Write data

Check code

Remarks: The check code is 16CRC with low byte ahead.

0x10 Continuous write data [HEX]

01

10

×× ××

××××

Address

Function code

Data

address

Register

number

××

×× ××

×× ××

 

Byte

number

Write data

Check

code

 

 

  • Format of register data

Address

Data name

Switch coefficient

Status

0

Temperature

0.1°C

R

1

DO

0.01mg/L

R

2

Saturability

0.1%DO

R

3

Sensor. null point

0.1%

R

4

Sensor. slope

0.1mV

R

5

Sensor. MV

0.1%S

R

6

System status. 01

Format 4*4bit 0xFFFF

R

7

System status.02

User command address

Format: 4*4bit 0xFFFF

R/W

Remarks:Data in each address is a 16-bit signed integer, the length is 2 bytes.

The real result=Register data * switch coefficient

Status:R=only read; R/W= read/write