Medical Distribution Transformers
Product Introduction
Used for power supply in important medical places (operating room, intensive care unit), when the direct leakage current of electrical equipment to the human heart is greater than 10uA, it will cause a micro-shock accident of the patient, and the leakage protection switch used in general general buildings, The action response is 1000uA level, which is far from meeting the needs of the medical field. Therefore, the important places in the medical field are required to use a local "isolated power system" for power supply.
Technical parameters
Rated capacity: 3.15/5/6.3/8/10kVA
Temperature rise: ≤80K (H grade)
Efficiency: ≥96%
Reference standard: GB/T 19212.16 "Special requirements and tests of isolation transformers for power supply in medical places"
Special electrical performance parameters: no-load current s3%In; surge current s12In(p); short-circuit voltage s4%Un
Working frequency: 50/60Hz
Cooling method: natural cooling
Industry standard products
Dimensions