Product Introduction
2200 series spectrophotometer is the ideal instrument for education and QC laboratories. They are simple to fit a curve by using your standard. They are widely used in colleges and quantitative analysis lab.
Features
Small fine design, space saving
The excellent optical and electric system
Low noise and Low stray light
Large LCD display (128*64 Dots), can display curve
High quality grating, detector and lamps
Data and Curve can be stored in real-time
Auto setting WL, auto Blank
Easy to change Pri-aligned lamps
Reinforced baseboard and bracket assures durability
Function
Photometric: T%, Abs
Quantitative: Standard Curve
System Utility
* WL Scan (Spectrum Scan) (PC)
* Time Scan (Kinetics) (PC)
* DNA/Protein Test (PC)
* Multi-WL Test (PC)
Specification
Model | EU-2200R |
Wavelength Range | 190-1100nm |
Spectral Bandwidth | 4.0nm/2.0nm |
Optical System | Split Beam, Blazed Holographic Grating (1200 lines/mm) |
Wavelength Accuracy | ±0.8nm |
Wavelength Repeatability | ≤0.2nm |
Wavelength Setting | Auto, Resolution 0.1nm |
Photometric Range | 0~200%T, -0.3~3A, 0~9999C |
Photometric Accuracy | ±0.002A (0~0.5A), ±0.004A (0.5~1A), ±0.3%T (0~99%T) |
Photometric Repeatability | ≤0.001A (0~0.5A), ≤0.002A (0.5~1A), ≤0.15%T (0~99%T) |
Stray Light | ≤0.05%T(220/360nm) |
Scan Speed | High, Medium, Low. Max.1000nm/minute |
Baseline Flatness | ±0.002A |
Stability | ±0.001A/h (500nm,0A) |
Noise | ≤0.2%T/3min (250/500nm,0%T); ≤0.5%T/3min (250/500nm,99%T) |
Sample Compartment | Accommodate 5-100mm Pathlength Cuvette |
Detector | Silicon Photodiode |
Lamps | Tungsten Lamp & Deuterium Lamp (Pre-aligned) |
Display | Graphic LCD (128x64 Dots) |
Keypad | 11-key Membrance Keypad |
Output Port | USB Port & RS232 Port |
Printer | Mini Serial Printer; PC Printer |
PC Software | Optional PC Scanning Software |
Power Requirements | AC 90-250V,50/60Hz |
Dimension | 431x318x195mm |
Weight | 10KG |