Product introduction
The LLX series is an air cooled, high frequency constant potential generator. It is dedicated to high technology applications like aerospace where specific requirements and speed of inspection are needed. These units are CP type and are controlled by a DC1 or Hand-XD and can be used in combination with digital detectors.
MAIN FEATURES
| KEY ADVANTAGES
| MOST COMMON APPLICATIONS
|
Extremely stable dose output for digital applications Constant potential high frequency generator Current up to 12mA Metal-ceramic tube Two focal spot available with Beryllium window Energy recovery module Safety first, x-ray start with 2 buttons
| Low weight Ergonomic rubber handle Extremely small head diameter Full automatic preheating process Low power requirement Cables with reduced size and weight Can be delivered with our 900g hand-XD including a Geiger-Müller detector RS232 serial link designed to be connected to digital detectors (ex. XPro Series) and computers Works on power generator CANON mil. metallic connectors
| Aeronautics, composites & plastics Building sites Digital applications (flat panels, CR or linear detectors) X-ray cabinets |
Specifications
| LLX110-DA-0 | LLX110-DA-1 | LLX160-DA-0 | LLX160-DA-1 |
| | |
| High voltage range | 5-110kV | 5-110kV | 5 - 160kV | 5 - 160kV |
| Tube current range | 0,5-12 mA | 0,5-12 mA | 0,5 - 12 mA | 0,5 - 12 mA |
| Maximum anode power dissipation | 600 | 900 | 600 | 900 |
| Beam angle | 60°x40° | 60°x40° | 60° x 40° | 60° x 40° |
| Focal spot size(IEC336) | 0.4x0.4 | 1.5x1.5 | 0.4x0.4 | 1.5x1.5 |
| Steel penetration at max (*) | 12 | 15 | 33 | 34 |
| Max recommended temp. | 50° | 50° | 50° | 50° |
| Temperature switch | 70° | 70° | 70° | 70° |
| Inherent filtration(mm) | 0,8 Be | 0,8 Be | 0,8Be | 0,8Be |
| Anode cooling type | Air | Air | Air | Air |
| Duty cycle | 99% | 99% | 99% | 99% |
| Mains | 180 - 260V | 180 - 260V | 180 - 260V | 180 - 260V |
| Weight (without rings) | 16 | 16 | 16 | 16 |
| Tubehead | Metal-ceramic | Metal-ceramic | Metal-ceramic | Metal-ceramic |
| Control unit | DC1 or Hand-XD | DC1 or Hand-XD | DC1 or Hand-XD | DC1 or Hand-XD |
| | | | |
| LLX200-DA-0 | LLX200-DA-1 | LLX200-PA-0 |
| | |
| High voltage range | 5 - 200kV | 5 - 200kV | 5 - 200kV |
| Tube current range | 0,5 - 12 mA | 0,5 - 12 mA | 0,5 - 12 mA |
| Maximum anode power dissipation | 600 | 900 | 600 |
| Beam angle | 60° x 40° | 60° x 40° | 360° x 40° |
| Focal spot size (IEC336) | 1.0 | 3.0 | 0.4x4 |
| Steel penetration at max (*) | 42 | 44 | 35 |
| Max recommended temp. | 50° | 50° | 50° |
| Temperature switch | 70° | 70° | 70° |
| Inherent filtration(mm) | 0,8Be | 0,8Be | 0.4Fe/Ni/Co |
| Anode cooling type | Air | Air | Air |
| Duty cycle | 99% | 99% | 99% |
| Mains | 180 - 260V | 180 - 260V | 180 - 260V |
| Weight (without rings) | 18.1 | 18.1 | 18.1 |
| Tubehead | Metal-ceramic | Metal-ceramic | Metal-ceramic |
| Control unit | DC1 or Hand-XD | DC1 or Hand-XD | DC1 or Hand-XD |
* FFD= 700mm/Film Carestream Industrex AA400/D=2.0/t=10min