Measurement Principle: Zirconia
Display Mode: 7-inch serial port capacitive screen
Measurement Range: 0~10/100/1000PPm/1.00%/25.00% O2
Measurement Accuracy: 0~1000PPm/1.00%/25.00%≤±1%FS
0~100PPm≤±2%FS
0~10PPm≤±5%FS
Resolution: 0.1PPm (trace)
0.01% (normal)
Repeatability: ≤±1.0%FS
Response Time: T90≤30S
Output Port: 8-channel 4-20mA.DC (Non-isolated output, load resistance< 1KΩ)
8-channel 0-10V.DC (Non-isolated output, load resistance> 10KΩ)
8-channel programmable alarm output (Dry contact type), maximum load AC220V/2A
Communication Mode: RS232 (default) or RS485
Power Supply: AC100~240V 50/60Hz, power dissipation<200VA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Temperature: 0℃~60℃
Sampling Mode: Extraction sampling
Sample Gas Flow: 400~600mL/min
Gas Inlet Pressure: -10kPa≤ relative pressure≤5kPa (when pressure <5kPa, please indicate specific negative pressure value when ordering)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or other inert gas
External Size: 235mm×400mm×422mm (H×W×D)
Hole-Cutting Size: 237mm×361mm (H×W) , embedded mounting
Sensor Service Life: >5 years (under normal conditions of use)
Gas Circuit Port: NPT 1/8 internal thread
Instrument Weight: Net weight 15.8Kg
Installation Method: Embedded mounting / counter mounting