Water Insoluble Substance Meter (Multi-factor Meter)
I. Equipment introduction (TC-6)
The device can be used to determine the filtration factor of polymer solution and the solution of filtered water or chemical agent.
The existing method for determining water-soluble polymer filtration factor used for oil displacement was developed based on the line standard SY/ t5862-2008. Gelman filter and 500ml measuring cylinder were mainly used to manually measure the outflow time per 100ml polymer solution (concentration 0.1%) with a stopwatch. In the past, the outflow time of most polyacrylamide solutions was between 0.5 and 3h, and the device could meet the measurement requirements at that time. However, with the development of eor technology, new products such as comb polymer and hydrophobic associative polymer have emerged. The performance change of the new polymer greatly lengthens the outflow time of the solution, and the total outflow time of 300ml solution can reach 20h. The night shift needs to be measured, which greatly increases the working strength. In addition, the use of stopwatch and cylinder measurement error is also large.
Principle of measurement: remove the measuring cylinder from the original filtration factor measurement and adopt a specially processed container with three groups of electrode probes, in which the liquid level conduction sensor monitors the liquid interface, and a single board machine can detect the 100ml, 200ml and 300ml liquid level and outflow time respectively, and time is measured by a precise time counter. In addition, a solenoid valve is added to the nitrogen gas source pipeline. When the 300ml outflow time is obtained, the single-board machine gives the instruction to close the solenoid valve, cut off the nitrogen gas source, and automatically end the measurement. Thus reduce labor intensity and reduce the workload of monitoring personnel.
Ii. Technical parameters
1. Air source pressure 1MPa, test pressure: 200KPa
2. Precision of time counter: 1 second
3. Plexiglass solution storage tank: 600ml
4. Filtration factor monitoring volume: 400ml
5. Parallel connection of 4 or 6 sets at the same time
6. Equipped with automatic feed and cleaning system
7. SMC pneumatic units are used for pipe and valve parts
8. Monitor and analyze the experimental data through PLC or monitoring software