Brief introduction of laboratory air supply system

Older laboratory in the use of gas in China are mostly put gas cylinders beside the point, and then install a pressure reducer in the cylinder at the exit, and then directly through copper tube or ptfe hose connection to the instrument, if a have multiple when using the equipment in the lab, there will be a lot of gas cylinders, at the same time, there are a lot of clutter line, at the same time there is no lack of these gases are flammable and explosive gas, toxic gas, corrosive gas, and the export of high pressure cylinders, make this way in real use process is full of danger. To solve these problems, modern laboratories have reformed the general atmosphere on which gases are to be used, which is known as a "gas network project" and a "concentrated gas supply system," in which gases are to be stored together and then to be transferred through metal or other piping to the point of use by means of differential pressure.

Its advantages mainly include the following:

1. Firstly, solve the placement problem of gas cylinders. A coffret if possible as far as possible in between the cylinder and the relatively independent laboratory room, if the lab in the same building, between the cylinder position is located in the stream of people less as far as possible and independent room, this way can make the cylinder and the personnel and equipment is completely isolated, even harmful gas leak, also direct damage will not occur.

Secondly, the problem of gas mixing. All the carrier gas cylinders are respectively concentrated in one cylinder room according to the gas properties (in fact, the most important is to store flammable and explosive gas (such as H2, natural gas, acetylene, etc.) and combustion gas (such as oxygen, chlorine gas) separately).

3. Again is to solve the problem of cylinder pressure. For each gas, multiple cylinders of gas can be connected in parallel and then transported to the point of use through an outlet for unified decompression. Because the cylinder room is relatively independent, the place where the pressure of the whole air circuit system is the highest is also the outlet of the cylinder. Therefore, this method reduces the possible risk of cylinder pressure to the cylinder room, and can minimize the damage to human body and the instrument.

4. Economy: a centralized cylinder room can save limited laboratory space, and it is not necessary to cut off the gas when replacing the cylinder to ensure the continuous supply of gas. Users need to manage fewer cylinders and pay less cylinder rent because all points of use using the same gas come from the same source. This method of supply ultimately reduces transportation costs, reduces the amount of residual gas returned to the gas company in the empty cylinders, and good cylinder management.

5. Higher utilization rate of laboratory environment: centralized pipeline supply system can place gas outlet at the point of use, so that the workplace can be designed more rationally.