SIN-TDS210-C Conductivity meter

Model
Nature of the Manufacturer
Other
Product Category
Place of Origin
City, Province
Overview
Category: Conductivity meter Views: 331 Applications: Thermal power, Chemical fertilizer, Environmental protection, Metallurgy, Pharmacy, Biochemistry, Food and water
Product Details





Measuring
Range
0.01 electrode:0.02 to 20.00us/cm
0.1 electrode:0.2 to 200.0us/cm

1.0 electrode:2-2000us/cm

10.0 electrode:20 to 20000us/cm
Accuracy
±1%FS
Stability
±1%FS/24H
Temperature
Compensation range:-10~130℃.NTC10K
Compensation
PT1000,accuracy:±0.5℃
2.4inch LCD
Relay out
2 ralay out,capacity:250VAC,3A
Signal output
4-20mA output,RS485 communictaion,compatible

with standad MODBUS-RTU
Power supply
AC220±10%,50Hz/60Hz
Dimensions
96*96*96.5mm(H*W*D)
Hole Size
92.5*92.5mm



Industry online pH meter controller


Install a 92.5 * 92.5 mounting hole on the instrument cabinet or mounting panel

The instrument into the mounting hole and then buckle on the butterfly, as shown below:



Installation of ph electrode


Schematic diagram of common installation method

①Side wall installation ②Flange mounted at the top ③Pipe installation ④Top installation ⑤Submersible installation ⑥Flow-through installation
The interface must be in 15°oblique angle, or it will affect the normal test and use of the electrode. We won’t be responsible for any results due to this.