HERMetic Sampler GTX Chem or Sampler GTN Chem

Honeywell
HERMetic Sampler GTX Chem or Sampler GTN Chem
GTX GTN
The HERMeticSampler GTX Chem and HERMetic Sampler GTN Chem are designedfor closed sampling of liquids or chemicals, which present a fire,health or airpollution hazard. The gas tight construction of these units avoids a pressurerelease from the tank and exposure to fumes during operation.
For the HERMetic Sampler GTX Chem the transfer ofthe liquid from the sampling bottle to a laboratory bottle occurs byoverpressuring the upper chamber of the sampler with a pump.
• After sampling, the liquid canbe transferred into a laboratory bottle by opening the transfer valve andactuating the pressure pump.
• The graduated tape permitschecking of sampling bottle height.
• The HERMetic Sampler GTX Chem isvery easy to clean and fully compatible with all kind of non corrosive andcorrosive liquids.
For the HERMetic Sampler GTN Chem, the samplecan be transferred under closed condition. This transfer guarantees theintegrity of the sample, since the liquid is never in contact with theatmosphere. A closed vapour recovery system dispatches the vapours back intothe tank during transfer of the liquid into the laboratory bottle.
• The HERMetic Sampler GTN Chemcan be purged with inert gas before and/or after sampling.
• The sampled liquid is never incontact with the atmosphere.
• The sampling height can bemeasured on the graduated tape.
Technical specifications:
HERMeticGTX Chem HERMetic GTN Chem
Maximum tankoverpressure : 0,3 bar 0,3 bar
Unit height: 800 mm801 mm
Weight : 5.3 kg 7.5 kg
Tape length : 30m /100 ft 30m / 100 ft
Capacity of samplingbottle : Approx. 0.5 l Approx. 0.5 l
Capacity of laboratorybottle : 0.47 l, 16 oz
Type of gaskets: FF KMFF KM
Materials: Stainlesssteel AISI 316, PTFE, PVDF
Hazardousenvironments approvals
ATEX II 1 G c IIB T6 II 1 G c IIB T6