Fiber Spectrometers for Industrial Control with High Sensitivities GLA639

Model
Nature of the Manufacturer
Other
Product Category
Spectrometers
Place of Origin
City, Province
Overview
Fiber Spectrometers for Industrial Control with High Sensitivities GLA639
Product Details

●Czerny-Turner optical structure with a grating, high end light detector and SMA905 fiber connector.

●Connects to a computer through USB, RS232 or other connector for industrial control, controlled and powered by the computer.

●Unique software features, automatic configuration of spectrometer, automatic reading of calibration coefficients, spectral acquisition and measurement.

●Spectral measurements including radiation, transmittance, absorbance and reflectivity.  

●Two output data formats, conventional Excel two-column format and square format, the latter can be used to check spectral data easily which is a unique design developed by us.

●The selected spectral data on the spectral curve are marked using different colors, including the cursors and the area boundary lines, allowing users to easily find data in a large number of spectral data.

●Cursors are used to find the wavelengths. The spectral width is found by moving the two boundary lines.

●Software logging allows the user to record experiment’s content such as the material analyzed, the operator and time etc.

●Color bars help users understand the corresponding wavelengths of the spectrum.

●Interfaces and data formats for industrial control, such as Linux, ARM and PLC etc.

●GLA639 fiber spectrometer includes UV bands spectrometer, VIS band spectrometer and NIR band spectrometer.  


●Excellent Consistency of mass production for industrial applications  

●Excellent consistency of thermal stability and vibration resistance

●Compact size and sleek appearance.


Items Specifications Remarks
Fiber Connector SMA905
Slit 10, 25, 50, 100µm
Wavelength Range 190-380nm/190-480nm Depends on the grating used
Spectral Resolution FWHM 0.27nm@Hg 253nm@25µm slit/
FWHM 0.4nm@Hg
253nm @25µm slit
Depends on wavelengths
Order Sorting Filter To eliminate higher order spectrum
Stray Light 0.1%@ 253nm Depends on wavelengths
Light Detector High end detector
Pixel Number 2048
Pixel Size 14μmx200μm
Detector Collecting Lens Yes
SNR 800:01:00 Room temperature
ADC Resolution 16 bits
Integration Time 1 ms-15min Depends on the detector used
Data Output Count vs. wavelength or pixel
Power Supply DC5V/380mA
Trigger Mode Trigger in or out with TTL
Communication USB o RS 232
Operation Software SPEC-GLA600
Application Programming Interface glaDevSys For users to develop software
Dimensions 100mmx65mmx36mm
Weight 285g
Mounting 4xM3 screw holes on the bottom
Operating Temperature -10-50℃
Storage Temperature -20-70℃


●Environmental inspection: inspection of water quality and air pollution etc.

●Chemical industry  

●Optical coating inspection

●Glass and transparent materials inspection

●Measuring light sources

●R&D in science and technology

●Customization