1950 Plus TOC Analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
TOC Analyzer
Place of Origin
City, Province
Overview
1950 Plus TOC Analyzer
Product Details
typical application
Used for surface water and tap water TOC content monitoring and TOC removal rate determination. Complies with standard method 5310C and EPA method 415.1.
 
Instrument characteristics
● Automatic dilution function, automatic correction, automatic cleaning
● Built-in intelligent software can automatically display TOC removal rate
● Two-channel sampling analysis
● Manual sample analysis
● Nitrogen or air can be used as carrier gas (through gas purifier)
● Automatic diagnosis, no water sample, no reagent, no gas alarm
● Comply with USEPA415.1, standard method 5310C, ISO8027 standard analysis method
 
Principle of analysis
The instrument adopts ultraviolet light / persulfate oxidation method.
The sample is first acidified to oxidize the total inorganic carbon (TIC) in the water to carbon dioxide; then the carrier gas is used to blow out the carbon dioxide; then persulfate is added to the water sample and oxidized with ultraviolet light (UV). TOC) is completely oxidized to carbon dioxide; then the carrier gas is used to bring the carbon dioxide into the infrared detector (NDIR) to detect the carbon dioxide concentration and convert the TOC concentration.
Scan the QR code to read on your phone
Technical index
Measuring range0 ~ 5/10 / 25mg / L
Response timeT90 ≤ 8 minutes
Accuracy± 2% of full scale
Lower detection limitAt 25 ℃, the measuring range is 0 ~ 5mg / L, ≤ 0.015mg / L
Injection conditions
0.15 ~ 6bar (2 ~ 87psig);
Flow rate, 25 ~ 200mL / min
Operating temperature5~40℃
Sample temperature2~70℃
Data communicationRS232 / RS485 serial port (optional)
Analog outputRS232 / RS485 serial port (optional)
Power requirements115 / 220V AC, 50 / 60Hz
Instrument specifications54kg ,981 × 675 × 220 mm