I. Introduction The T ransformer oil dissolved gas analyzer is suitable for the determination of the dissolved gas content in the insulating oil of the power system. The total analysis of the dissolved gas content in the insulating oil can be completed by one injection, and the minimum detection...
Product Details

I. Introduction

The Transformer oil dissolved gas analyzer is suitable for the determination of the dissolved gas content in the insulating oil of the power system. The total analysis of the dissolved gas content in the insulating oil can be completed by one injection, and the minimum detection concentration of acetylene is 0.1ppm.The instrument is equipped with large screen LCD display interface, menu type English operation, intuitive display and convenient operation.The instrument adopts double column parallel shunt system, equipped with thermal conductivity detector, double hydrogen flame detector and nickel catalyst converter, which can complete the analysis of H2, CO,CO2,CH4, C2H2, C2H4, C2H6,O2, C3H8, N2 in the transformer oil

1

II. Main function and configuration

Gas analyzing

it can detect 10 gas of transformer oil: H2, CO,CO2,CH4, C2H2, C2H4, C2H6,O2, C3H8, N2 in the transformer oil.

Item

Name

Performance description

Quantity

1) HM-GC20 DGA

1.

Host

Specification

Remark

1 set

2.

Hydrogen flame ionization detector

FID

Detection limit ≤1Pg

Baseline noise ≤1PA

Baseline drift ≤1PA/30min

Linearity ≥106

/Automatic ignition function/Wide range amplifier


2 set

3.

Thermal conductivity detector

TCD

Sensitivity S≥6000mV•ml/mg(n-hexadecane / isooctane)

Baseline noise ≤20μV

Baseline drift ≤30μV/30min

Linearity ≥104


1 set

4.

Nickel catalyst converterBuilt-in nickel catalyst column

Converter furnace temperature control range0400℃

Methylation conversion temperature350℃380℃

Temperature control modePlatinum resistance100Ω);

Plus potential power90W×2Internal heat type);

Methane conversion rate≥99%


1 set

5.

sample injector

Fill column systemA two-column shunt system was used


2set

6.

Chromatographic analysis column

GDX-502 specifications4m*3mm


1pc

7.

Chromatographic analysis column

5A molecular sieve specifications1m*3mm


1pc

8.

Chromatographic analysis column

TDX-01 specifications1m*3mm


1pc

9.

work station

Three channels of the reverse control version of the transformer oil special workstation


1set

10.

spare part

sample needle, sampler, extension needle, pressure reduction valve joint, sample pad, sealant cap, two-way needle, etc


1set

2)Other configurations


1.

Generators

ZHHK-3 hydrogen gas generator

hydrogen purity 99.999%

Hydrogen flow0300ml/min

Traffic displayLED digital display

working pressure 00.5MPaSet 0.3MPa at delivery

Power Supply Voltage:~220V ±10% 50Hz

consumed power130W


1pcs

ZHAK-2 Air generator

output flow 02000ml/min

output pressure00.4MPa

system pressure0.550.75MPa

Stable pressure accuracy0.003 MPa

Work noise:<40Dbd(A)

Power Supply Voltage:~220V ±10% 50Hz

consumed power150W


1pcs

2.

cylinder

argon99.999%Cylinder + decompression valve

By buyer

1set

3.

Standard gas

H2CH4C2H2C2H4C2H6C3H8 N2 O2COCO2 Gas component standard gas (including pressure reducing valve) 8L aluminum bottle


1bottle

4.

Deoxygenation tube

Adadsorption of trace oxygen in carrying gas


1set

5.

Multi-function degassing oscillator

1.Large-screen LCD display, English menu input, using microcomputer control, PID self-setting and fixed temperature control mode, the temperature control is more accurate

2.Mechanical transmission is fatigue resistant, vibration resistant, maintenance free and low noise

3.There is a plug thermometer socket for the operator to observe and compare

4.Small size, light weight, elegant appearance

5.Eight pcs 50 ml or 100ml syringes at the same time

6.Four pcs 250ml triangular bottles can be placed at the same time

7.temperature control indoor -50℃±0.3℃

50℃~100℃±0.5℃

8.time controlOscillation time: 0-59 points is optional

Still time: 0-59 points

Select the oscillation frequency: 275 times / minute ± 3 times


1set

6.

Computers and printers

Lenovo desktop computer 19-inch LCD screen


1set

III.Feature

HM-GC20 Host Feature

★Adopt advanced Gigabit / Gigabit Ethernet communication interface, built-in IP protocol stack, enables the instrument to achieve long-distance data transmission; convenient laboratory setup, simplify laboratory configuration, convenient analysis data management;

★The instrument internal design of 3 independent connection IP addresses, can be connected to the working computer (laboratory site), in charge of the computer (such as quality inspection department, production department, etc.), and senior management computer (such as the Environmental Protection Bureau, technical supervision Bureau, etc.), can monitor the operation of the instrument, and analyze the data results in real time when needed;

★The special workstation equipped with the instrument can support multiple chromatograph access at the same time, realize data processing and instrument reverse control, simplify document management;★The instrument can be connected to the manufacturer through the Internet to realize remote diagnosis, remote update, etc. (user Settings);

★The instrument is equipped with an 8-inch color LCD touch screen, support hot plug, can be used as a handheld controller

★The instrument works in parallel to make it more stable and reliable, with various optional high-performance detectors, such as FID, TCD, ECD, FPD and NPD, to meet the analysis requirements of complex samples.

★The instrument adopts the modular structure design, and the later maintenance is simple and convenient.★The new microcomputer temperature control system, high temperature control accuracy, superior reliability and anti-interference performance; with eight completely independent temperature control output, can achieve 20 order program heating, with column box automatic rear door opening system, near room temperature control ability is improved, liter / cooling speed is faster;

★The instrument configuration of advanced electronic flow control unit (EFC), electronic pressure control unit (EPC) to realize the digital control of the gas path, greatly improve the stability of the instrument and the reproducibility of the analysis results;

★The chromatograph has a low-noise, high-resolution 24-bit AD circuit, and has a baseline storage, baseline deduction function.

★The standard workstation is suitable for WinXP, Win2000, Win7, Win8, Win10 and other operating systems.★The chromatograph microcomputer system with fully independent intellectual property rights has the standard protocol of MODBUS / TCP, and can connect with the DCS system.

Host Technical

●Operation display: a 5.7-inch color LCD touch screen

●Temperature control area: 8 channels

●Temperature control range: 4℃ ~450℃ above room temperature, increment: 1℃, precision: ± 0.1℃

●Program heating order: 20

●Range lift rate: 0.1~60℃ / min

●Gas path control: full electronic pressure (EPC), electronic flow rate (EFC) control

●Range: 0~100Psi (pressure); 0 ~ 1000 mL/min (flow)

●Resolution: 0.01Psi (pressure); 0.01 mL/min (flow)

●External event: 8 channels; auxiliary control output 2 channels

●Sampler type: filling column inlet sample, capillary sample inlet sample, six-way valve gas inlet sample, automatic sample inlet device

●Number of detectors: 3 (maximum); the FID, TCD, ECD, FPD, and NPD are optional

●Start the incoming sample: manual, automatically optional

●Communication Interface: Ethernet: IEEE802.3

Generators technical

Hydrogen flame ionization detector FID

Detection limit ≤1Pg

Baseline noise ≤1PA

Baseline drift ≤1PA/30min

Linearity ≥106

Thermal conductivity detectorTCD

Sensitivity S≥6000mV•ml/mg(n-hexadecane / isooctane)

Baseline noise ≤20μV

Baseline drift ≤30μV/30min

Linearity ≥104

Electron capture detector (ECD)

Detection limits ≤1×10-13g/mlPropyl hexa-hexa-isooctane solution

Baseline noise ≤0.03mV

Baseline drift ≤0.2mV/30min

Linearity103

●radioactive source 63Ni

flame photometric detector (FPD)

Detection limits (S)≤5×10-11g/s

(P)≤1×10-12g/sMethyl parathion - absolute ethanol solution

●baseline noise≤3×10-13A

● baseline drift ≤2×10-12A/30min

● linearity rangeto sulphur≥102to phosphorus≥103

Backcontrol chromatography workstation

The HM-GC-20 gas chromatography, different from a general gas chromatograph, requires a 10 / 100M adaptive network environment (Internet access services when remote data transmission is needed over the Internet) and a computer already equipped with English Windows2000, XP, win7, win8 operating systems.

Below is the working interface.

1659495622587


Technical indicator

Analytical detection limits of dissolved gases(uL/L)

H2

O2

CO

CO2

CH4

C2H4

C2H6

C2H2

1

1

2

2

0.05

0.05

0.05

0.05

───────────────────────

Items Reserved time Gas potency peak area

───────────────────────

1 0.455 H2 100 36392

2 2.040 CO 70 1529926

3 12.922 CO2 300 8835319

4 17.572 CH4 10 213463

5 19.316 C2H6 10 354775

6 19.823 C2H4 10 402530

7 20.435 C2H2 5 192036

───────────────────────

Total hydrocarbon concentration 35

Three-ratio coding100