Welcome Customer !

Membership

Jiuyin Science and Technology Chengdu Co., Ltd.
Free Member ·1YRS
ChengDou,SiChuan
Manufacturer

Main Products: Oxygen Nitrogen Analyzer (Oxygen Nitrogen Hydrogen Analyzer)|Automobile Exhaust Gas Analyzer

Xwboo>Products

Jiuyin Science and Technology Chengdu Co., Ltd.Free·1YRS

  • E-mail

    sales@meas-elec.com,service@meas-elec.com

  • Phone

    028-85044052/028-85044053

  • Address

    4th floor, unit 1, building 4, District D, West Zhigu, 66, Wuxing 4th Road, Wuhou District, Chengdu City, Sichuan Province

Contact Now

HT-LA521 High-Content Oxygen Analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
High oxygen analyzer
Place of Origin
City, Province
Overview
HT-LA521 High-Content Oxygen Analyzer
Product Details

    Measurement Principle: Ion current (limiting current)

    Display Mode: 320×240 graphic dot matrix color LCD

    Measurement Range: 10.00%~98.00% O2

    Measurement Accuracy: ≤±1.0%FS

    Resolution: 0.01%

    Repeatability: ≤±1.0%FS

    Response Time: T90≤30S

    Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)

    0-10V.DC (Non-isolated output, load resistance>10KΩ)

    2 ways of programmable alarm output (Dry contact type, maximum load 220V/2A)

    Communication Mode: RS232 (default) or RS485

    Power Supply: AC85~264V 50/60Hz, power dissipation <25VA

    Environment Temperature: -10℃~+50℃

    Storage Environment Humidity: <90%RH, non-condensing

    Operation Environment Humidity: <100%RH, non-condensing

    Sample Gas Temperature: 0℃~50℃

    Sampling Mode: Gas in

    Sample Gas Flow: 400~600mL/min

    Gas Inlet Pressure: 5Kpa≤ relative pressure≤200Kpa (relief valve is recommended if the pressure >200Kpa)

    Gas Outlet Pressure: Free venting (under safe conditions)

    Background Gas: N2 or other inert gas

    External Size: 155mm×155mm×255mm (H×W×D)

    Hole-Cutting Size: 135mm×135mm (H×W)

    Sensor Service Life: short life type >4 years, long life type >6 years (under normal conditions of use)

    Gas Circuit Port: NPT 1/8 internal thread

    Installation Method: Embedded mounting

    a. Friendly communication menu allows an intuitive and convenient operation;

    b. 320×240 graphic dot-matrix color LCD display is clear to see;

    c. The sensor consumes no service life when power is off, and could enter protection mode automatically when range is beyond the required limit with sensor’s automatic protection function to extend sensor’s service life;

    d. Automatic temperature and pressure compensation could eliminate the impact of environment temperature and pressure changes on the measurement accuracy;

    e. No calibration gas is needed and the instrument is not affected by the oxygen content in the operation environment;

    f. The instrument allows multi-point fractional liner calibration, meeting the need for an accurate measurement of oxygen content among the full range;

    g. RS232 (default) or RS485 digital communication port can be selected according to clients’ requirement, to realize unidirectional or bidirectional communication with computer or other digital communication facilities.

    It is widely used in the real-time detection and analysis of high concentration oxygen in large air separation for high purity oxygen, medical and industrial oxygenerator, smelting industry, medical and health care, petrochemical industry, environmental protection, electronic power and other industries.

    Measurement range

    Measured gas pressure

    Background gas composition