Measurement Principle: Ion current (limiting current)
Display Mode: 128×64 dot matrix LCD
Measurement Range: 10.00%~99.99% O2
Measurement Accuracy: ≤±1.0%FS
Resolution: 0.01%
Repeatability: ≤±1.0%FS
Response Time: T90≤30S
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
0-10V.DC (Non-isolated output, load resistance>10KΩ)
1 way of programmable alarm output (Dry contact type, maximum load 220V/2A)
Communication Mode: RS232 (default) or RS485
Power Supply: AC170~264V 50/60Hz, power dissipation <20VA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Temperature: 0℃~50℃
Sampling Mode: Gas in
Sample Gas Flow: 400~600mL/min
Gas Inlet Pressure: 5Kpa≤ relative pressure≤200Kpa (relief valve is recommended if the pressure >200Kpa)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or other inert gas
External Size: 155mm×155mm×296mm (H×W×D)
Hole-Cutting Size: 135mm×135mm (H×W)
Sensor Service Life: short life type >4 years, long life type >6 years (under normal conditions of use)
Gas Circuit Port: NPT 1/8 internal thread
Installation Method: Embedded mounting