Measurement Principle: Ion current (limiting current)
Display Mode: External display module (optional)
Measurement Range: 10.00%~98.00% O2
Measurement Accuracy: ≤±1.5%FS
Resolution: 0.01%
Repeatability: ≤±1.0%FS
Response Time: T90≤30S
Output Port: 4-20mA.DC (Non-isolated output, load resistance<Ω)
1 way of programmable alarm output (Dry contact type, maximum load 220V/2A)
Display module port
Power Supply: DC24V±10%, 1A
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Temperature: 0℃~50℃
Sampling Mode: Gas in
Gas Inlet Pressure: Stable pressure (control the pressure within the recommended limits to make sure the flow is within 400~600mL/min)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or other inert gas
Size: 29mm×75mm×162mm (H×W×D)
Sensor Service Life: >4 years (under normal conditions of use)
Gas Circuit Port: Φ6 plastic tube fitting
Installation Method: Wall mounting