On-line digital turbidity sensor (economic model) BH-485-TB02
Turbidity sensor is based on infrared scattering light technology, that is, the infrared light emitted by the light source will be scattered when passing through the measured sample during transmission, and its scattered light intensity is proportional to turbidity. The BH-485-TB02 turbidity sensor is equipped with a scattered light receiver in the direction of 90, and the turbidity value is obtained by analyzing the intensity of the scattered light.

Instrument features:
infrared scattering light technology can eliminate the influence of sample color; The automatic cleaning function of cleaning brush can be selected, which greatly reduces the maintenance of sensor; Digital sensor with strong anti-interference ability and long transmission distance; Standard digital signal output can realize integration and networking with other devices without controller
Typical application:
on-line monitoring of turbidity in each process of sewage treatment plant; On-line turbidity monitoring of water and wastewater treatment processes in various industrial production processes
measuring range | turbidity:0.01-3000 NTU |
measurement accuracy | Less than 5% of the measured value, or ±0.5NTU, whichever is greater |
repetitiveness
| ±3% |
resolution ratio
| 0.01NTU , 0.1NTU ,Depending on the range |
pressure range | ≤0.2Mpa |
calibrate | Calibration of standard solution and water sample |
Main material of sensor
| Fuselage: SUS316L;; Upper and lower covers: PPS+ glass fiber, cable: PUR |
Power Supply | 9~36VDC |
protocol | MODBUS RS485 |
Storage temperature | -15到50℃ |
Working temperature | 0 to 45℃ (no ice) |
measure | Sensor: diameter 49mm* length 230.5mm |
weight | sensor:0.8KG |
the protection grades | sensor:IP68/NEMA6P |
Cable length | Standard 10-meter cable can be extended to 100 meters |