Y09-AD310 Aerosol diluter
Feature:
1. It operates between a 1 cubic foot / minute particle counter and an aerosol sampling tube.
2. The dilution ratio can be 10:1 or 100:1, and the diluted aerosol types include PAO, DOP, Ondina and so on.
3. The differential pressure monitoring system is mainly used to monitor the pressure drop at the inlet and outlet of the diluter.
4. The flow control and feedback system is used to monitor the work flow and feedback function of the diluter.
5. Professional team development and production, with patented technology, quality assurance

Tag: Aerosol diluter、Aerosol generator、Cleanroom、 Dust Particle Counter、Tabletop aerosol diluter
| Sampling | 28.3L/min (Deviation≤±5%) |
| Aerosol | PAO-4, DOP, Ondina, Mineral Oil, Paraffin |
| Particle size range | 0.1μm, 0.2μm, 0.3μm, 0.5μm |
| Dilution ratio | 10times, 50times, 100times |
| Inlet and outlet diameter | 11mm |
| Power | No need (to use a counter or photometer suction) |
| Repeatability | Standard Deviation≤10% |
| Size | 333X303X151.5 (mm) |
| Weight | 7KG |
| Material | SUS316L |
| Reference Standard | ISO 14644-1:2015; Y09 Serial laser dust particle counter enterprise standard; Q/320500GZY001-2016; GB/T 6167-2007 Test method for performance of dust particle counter; JJF-1190-2008Calibration specification for dust particle counter measurement; |