Welcome Customer !

Membership

Beijing Zetron Technology Co., Ltd
Free Member ·1YRS
Other

Main Products: 气体检测仪

Xwboo>Products
Product Categories

Beijing Zetron Technology Co., LtdFree·1YRS

  • E-mail

    info@bjzetron.com

  • Phone

  • Address

Contact Now

Micro Flow Control Portable Infrared Gas Analyzer Gasboard-3800P

Model
Nature of the Manufacturer
Other
Product Category
Place of Origin
City, Province
Overview
It adopts international leading UV DOAS technology, NDIR technology and long-life ECD technology, and is equipped with an integrated and automatic sampling pretreatment unit and control unit, Gasboard-9081 is suitable to simultaneously measure NO, SO2,O2, O2, CO and CO2 concentration in flue gas. It is an on-line monitoring device dedicated to boiler emission and energy efficiency control, which meets the emission regulatory requirements of national and local environmental protection departments.
Product Details

Details

• Integrate heated sampling probe and sampling hose.
• Built-in heated filter, prevent sampling loss.
• Max working temperature 200℃.
• Smart design, safe and reliable.
• Filter element: 2um
• Length: 3m
• Power supply: 220V
• Power consumption: 330W
• environmental temperature: -40℃-140℃

Gas analyzer features:
• Innovative micro-flow NDIR technology.
• Measure SO2, NO,CO,CO2,O2 and optional temperature, pressure and flow.
• Less interference of H2O for SO2 and NO measurement.
• Built-in zero pump and auto zero calibration with fresh air.
• Optional printer.
• Calculation of Lambda and combustion efficiency.
• Touch screen.
• USB output.

Gas conditioning system features:
• Efficient peltier dehumidifiers for humidity measurement of low concentrations.
• PID temperature control.
• Needle valve flow control.
• Built-in sampling pump and peristaltic pump.

Parameter

General Parameter
MeasurementSO2, NO, CO, CO2, O2
TechnologySO2, NO, CO, CO2 (Micro-flow NDIR)
O2 (Electrochemical)
RangeSO2: 0-2000ppm
NO: 0-2000ppm
CO: 0-5000ppm
CO2: 0-25%
O2: 0-25%
(Note: measurement range can be customized.)
ResolutionSO2, NO, CO: 1ppm
CO2, O2: 0.01%
PrecisionSO2, NO, CO, CO2≤±2%,O2≤±3%
RepeatabilitySO2, NO, CO, CO2, O2 ≤1%
Zero/Span driftSO2, NO, CO, CO2≤±2%,O2≤±3%
Response time (T90)<>
Warm-up time30 minutes
Communication InterfaceRS232
Work temperature0-50 ℃
Relaive humidity5-85%
Ambient prestsure86-108kPa
Power supplyAC 220V 50Hz±1Hz

Application

Combustion and emission control research or certification of Continuous Emission Monitoring System (CEMS), combustion efficiency monitoring for various industrial furnaces or refuse incinerator and boiler.