Carbon Monoxide 5-100ppm
Gas Detection Tube
Overview
The gas detector tube is a tool for rapidly measuring the concentration of harmful gases in the air. It is a fine glass tube filled with a certain amount of detection agent and fixed with materials, and then the two ends of the heating melt closed. The detection agent is generally made of silica gel, activated alumina, glass particles and so on, so that the detection tube has the advantages of simple and rapid determination of pollutant concentration
Product Paramenters
Working principle :
All kinds of gas detection tube is in a fixed finite length of the inner diameter of the glass tube, filled with a certain amount of detection agent (indicating powder), with the plug to be fixed, and then the two ends of the glass tube sealed processing. The detection agent is a kind of substance that can react with the substance to be measured and can change the color of the chemical reagent adsorbed on the surface of the solid carrier particle. The choice of chemical reagent and its chemical concentration ratio on the carrier determine the substance composition and range of the detection tube.
Manual sampler is required for the use of detection tube
Carbon monoxide detection tube including
5-100mg/m3,
10-200mg/m3,
20-500 mg/m3,
50-1000 mg/m3,
200-5000 mg/m3
The length of the test tube is 140mm, and the inner diameter is 2.4~3.5
Detection method
1. Check whether the vacuum sampler handle is at the zero position, and whether the red dot of the sampling mark of the handle is aligned with the sampling mark line of the sampler barrel. If not, restore the zero position and align the red dot of the sampling mark of the handle with the sampling mark line of the sampler barrel.
2. Insert the sealing of the gas detection tube into the round hole on the front side of the sampler before measurement, and skimmer off the sealing head of the gas outlet end and air inlet end of the detection tube in turn. Check the opening of the head. It should be opened, otherwise it should be skimmed off again.
3. Insert the outlet end of the gas detection tube into the air inlet of the sampler and plug it tightly.
4. The inlet of the measuring pipe will receive the measured air, pull the sampler handle to the second gear, pull it forward and backward once, observe the stopwatch, control the intake time for 2 ~ 3 minutes (0.6 ~ 0.9mm/s), and sample 100ml in total.
5. When the indicator discoloration in the detection tube (blue → white) stops, remove the detection tube and read out the ozone concentration from the white column to the scale.
6. Restore the vacuum sampler handle to zero.
Product Feature
1. On-site measurements can be performed easily by anybody, anytime and anywhere, in a short time
2. Easy-to-check direct reading possible due to calibrated scale directly printed on the tube
3. An extensive measuring range can be obtained by adjusting the sampling volume
4. In order to assure high precision, each production lot is tested and calibrated independently,
Each detector tube has its own quality control number printed on the tube
5. Long shelf-life with excellent long-term stability
Manual sampler is required for the use of detection tube
Carbon monoxide detection tube including
5-100mg/m3,
10-200mg/m3,
20-500 mg/m3,
50-1000 mg/m3,
200-5000 mg/m3
The length of the test tube is 140mm, and the inner diameter is 2.4~3.5