Measurement Principle: Ceramic humidity
Display Mode: 4 digit high-light LED display, display resolution 0.01%
Measurement Range: 0~40.00%VOL
Measurement Accuracy: ≤±2.0%FS
Repeatability: ≤±1.0%FS
Response Time: T90<15S
Output Port: 4-20mA.DC (Non-isolated output, load resistance<400Ω)
0-5V.DC (Non-isolated output, load resistance>10KΩ)
1 way of programmable alarm output (Dry contact type, maximum load 220VAC/2A)
Communication Mode: RS485 (default) or RS232
Power supply: AC220V±10% 50/60Hz, power dissipation <40VA
Instrument Operation Temperature: -10℃~+50℃
Probe Operation Temperature: Normal type: 0~200℃, high-temperature type: 0~400℃,special type: 0~650℃
Sampling Mode: In-site sampling
Sample Gas Composition: No flammable gas and strong-corrosive gas
Protection Level: Probe material: 316L, header protection level: IP65
Probe Length: Normal length 0.5m、1.0m、1.2m, other length is customizable (≤1.8m)
Service Life: Instrument >36 months, sensor >24 months (under normal conditions of use)
Instrument Weight: Net weight 7.1Kg (1.0m)
Installation Method: plug-in, standard DN65 flanged mounting