Measurement Principle: Electric capacity
Display Mode: 4 digit high-light LED display, display resolution 0.01%
Measurement Range: Volume absolute humidity: 0~40.0%VOL
Relative humidity: 0~99%RH
High-temperature dew point: -20℃~+100℃
(Default: Volume absolute humidity. Only one mode can be displayed and cannot be changed once leaving the factory. Please indicate the display mode you need when ordering.)
Measurement Accuracy: ≤±1.5%FS comprehensive accuracy, high-temperature dew point ≤±0.5℃
Repeatability: ≤±1.0%FS
Response Time: T90<30S
Output Port: 4-20mA.DC (Non-isolated output, load resistance<1KΩ)
0-10V.DC (Non-isolated output, load resistance>10KΩ)
1 way of programmable alarm output (Dry contact type, maximum load 220VAC/2A)
Communication Mode: RS485 (default) or RS232
Power Supply: AC100~240V 50/60Hz, power dissipation <80VA
Sampling Mode: Gas in
Sample Gas Temperature: 0~170℃
Gas Circuit Port: NPT 1/4-Φ6 stainless-steel fitting (316L)
Size: 110mm×220mm×309mm (H×W×D)
Install Hole Size: 168mm×205mm (H×W)
Sensor Service Life: >24 months (under normal conditions of use)
Instrument Weight: Net Weight 3.5Kg
Installation Method: Wall mounting